Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 12 of 12
BLEASE QUALITY DOCUMENT
BQD 083.V2
25/5/94
4.18 Parameter checking
Check that the ventilator monitored Vte &Vti are within tolerance at the following settings
Set Tv 0.2, BPM 25, I:E 1:2 Measured Tv = Set Tv +/- 0.03
Set Tv 0.5, BPM 12, I:E 1:2 Measured Tv = Set Tv +/- 0.05
Set Tv 0.8, BPM 10, I:E 1:2 Measured Tv = Set Tv +/- 0.05
Set Tv 1.0, BPM 6, I:E 1:1.0 Measured Tv = Set Tv +/- 0.05
Also confirm that the ventilator monitored peak pressure = the externally monitored peak
pressure +/- 2cmH
2
O.
4.19 Assembly Completion and burn in
At this point the case wrap should be put on the unit and the unit should be put on burn
in for a further 24 hours before following final test procedure 129TP000.