Technical Manual
3-12
Range/Alternatives Explanation and
<default value> result of alternatives.
Operator Lock
Off
On
<Off>
Off: Operator lock is not activated.
On: Operator lock is activated, preventing unauthorized
access to the instrument.
See chapter ‘Operation – Security locks'.
Operator Code
Range:
1 - 9999
<1937>
Defines the valid code for Operator lock.
If ‘Set-up lock’ (see below) is ‘On’ this code
will not give access to ‘Parameter Set-up’.
Note: this parameter is only shown if ‘Operator Lock’
is set to ‘On’
Set-up Lock
Off
On
<Off>
Off: Set-up lock is not activated.
On: Set-up lock is activated, preventing unauthorized
access to the instrument.
See chapter ‘Operation – Security locks'.
Set-up Code
Range:
1 - 9999
<1937>
Defines the valid code for Set-up lock.
If ‘Operator lock’ (see above) is ‘On’ this code will still
give access to all menus in the Main menu.
Note: this parameter is only shown if ‘Set-up Lock ‘
is set to ‘On’
Warm Up Time
Range:
0 - 200
Unit: min
<0>
Defines the delay time in minutes from power up until
the weight presentation has full accuracy.
Indicated in the display with the text ‘Warming up!’.