Medical Electrical Equipment needs
special precautions regarding EMC
and needs to be installed and put
into service according to the EMC
information provided in the following.
Portable and mobile RF
communication equipment (e.g. cell
phones) can affect Medical
Electrical Equipment. The use of
accessories and cables other than
those specified may result in
increased emissions or decreased
immunity of the unit.
Recommended separation distances between
Portable and mobile RF communications equipment and the ME equipment#
The Wireless Blood Pressure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Wireless Blood Pressure Monitor can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Wireless Blood
Pressure Monitor as recommended below , according to the maximum output power of the communications equipment.