E-20A-H-O / E-20AJ-H-O User Manual V1.3
4 / 34
5. Specifications
Device Size
Dimensions: 170 mm × 196 mm × 118 mm, or 290 mm × 196 mm × 134 mm (with the
humidifier)
Weight:1.5 kg, or 2.5 kg (with the humidifier)
Product Use, Transport and Storage
Operation Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing 15% to 93% Non-condensing
Atmospheric Pressure: 760 ~ 1060 hPa 760 ~ 1060 hPa
Mode of Operation
Continuous
Work Mode
CPAP, Auto
SD Card
With a capacity≥2 G, the SD card can record patient data and fault information. Furthermore,
the language pack stored on the SD card enables you to change the language of the device.
AC Power Consumption
100 ~ 240 V AC, 50 / 60 Hz, 2.0 A max
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
4 to 20 hPa (in 0.5 hPa increments), ≤30 hPa under single fault conditions.
Pressure Display Accuracy
± (0.5 hPa+4%)
Pressure Stability
4 to 20 hPa (± 1 hPa)
Ramp
The ramp time ranges from 0 to 60 minutes.
Sound Pressure Level
<30 dB, when the device is working at the pressure of 10 hPa.