EasyManua.ls Logo

BodiTrak LT System - Care of the Boditrak Mat; Cleaning the Boditrak Mat

Default Icon
14 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SH042-1 User Manual for BodiTrak LT V1.5, 25 Jan 2019
14
BodiTrak Technical Support: 800-847-3157
6
CARE OF THE BODITRAK MAT
Do not pull on or carry the mat by the USB cable.
Place the mat gently on surfaces by using the corners.
Do not pivot on the mat.
Do not pull on the edges of the mat while it is under a client.
The covering on the pressure sensing mat should not be in direct contact with broken
skin.
Do not pinch the mat between moving components of a support surface.
Do not fold the mat
Store the mat flat or in the original packaging by rolling the mat around the foam tube
and storing in the shipping cylinder.
If your mat has a ferrite installed on the USB cable do not remove it.
It is the operator's responsibility to ensure all cables are in good condition; inspect all
cables for nicks or abrasions prior to each use.
7
CLEANING THE BODITRAK MAT
BodiTrak pressure sensing mats are typically covered in polyurethane coated nylon/Lycra
stretch fabric.
Unplug the mat before cleaning.
For daily cleaning, we recommend wiping down with a damp, not wet cloth using a mild
soap and water solution.
Mild alcohol based cleaners and cleaners containing no more than 10% bleach in non-
wetting amounts may also be used.
Other cleaners / disinfectants sold under the following names may also be used:
Viraguard, Cavicide, Vitre Tb, and SeptFx.
Undiluted bleach and Hydrogen Peroxide are not recommended.
Cleaning products or disinfectants must be thoroughly wiped off after using them.
Allow the mat to dry thoroughly before use or storage.