Prior to any fault finding, please ensure all connections to the chiller are sound and that the incoming supply
is turned on. Also ensure that all electrical connections to the chiller and in the chiller are secure and in
good condition, the power is on and that the chiller has had adequate time to reach operating temperature
NOTE: Isolate from mains before removing any panels
Check connections to water supply.
Check water supply is turned on.
Check system for blockages.
Check thermostat/temperature probe is
correctly located into the bath probe well.
Check the agitator is running. If supply
voltage is present renew agitator
assembly.
If agitator is running with no water
agitation check agitation blades.
Insufficient air flow
through the fridge.
Check that the condenser is not blocked.
Check for blockages and obstructions to
ventilation grills.
Check supply to cooling fans.
If supply present replace fans.
If supply not present check connections,
thermostat, high side protection (digital
thermostat only) and fuse.
Check supply to Compressor.
If supply present return for repair.
If supply not present check connections,
thermostat, high side protection (digital
thermostat only) and fuse.
If compressor & fan are running and there
is no cooling, return for repair.