Guidance and manufacturer’s declaration - electromagnetic immunity
The Rotablator™ System is intended for use in the electromagnetic environment specified below. The customer or the user of the Rotablator System should assure
that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 V
3 V/m
Portable and mobile RF communications equipment should be used no closer to
any part of the Rotablator System, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2 P
d = 1,2 P 80 MHz to 800 MHz
d = 2,3 P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey
a
, should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the Rotablator System is used exceeds the applicable RF
compliance level above, the Rotablator System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Rotablator System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the Rotablator System
The Rotablator System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
Rotablator System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Rotablator System as recommended below, according to the maximum output power of the communications equipment.
Rate Maximum Output Power of
Transmitter
W
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d = 1,2 P
80 MHz to 800 MHz
d = 1,2 P
800 MHz to 2,5 GHz
d = 2,3 P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Warranty
Boston Scientific Corporation (BSC) warrants that reasonable care has been
used in the design and manufacture of this instrument. This warranty is in lieu
of and excludes all other warranties not expressly set forth herein, whether
express or implied by operation of law or otherwise, including, but not
limited to, any implied warranties of merchantability or fitness for a
particular purpose. Handling, storage, cleaning and sterilization of this
instrument as well as other factors relating to the patient, diagnosis, treatment,
surgical procedures and other matters beyond BSC’s control directly affect the
instrument and the results obtained from its use. BSC’s obligation under this
warranty is limited to the repair or replacement of this instrument and BSC
shall not be liable for any incidental or consequential loss, damage or expense
directly or indirectly arising from the use of this instrument. BSC neither
assumes, nor authorizes any other person to assume for it, any other or
additional liability or responsibility in connection with this instrument.