EasyManua.ls Logo

Boxym bpw1 - LCD Error Message and Cause Table

Default Icon
23 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V1.0
20
atmospheric
requirements, and
non-corrosive gases
and ventilated
room.
Security Type
Internal power supply, BF type application equipment
Duration of use
5 years
. Blood pressure classification instructions
Table 1 Blood pressure classification instructions
LCD
Systolic Pressure
Diastolic Pressure
Serious
≥180
≥110
Moderate
160179
100109
Mild
140159
9099
Little higher
130139
8589
Normal
120129
8084
Most suitable
≤119
≤79
. LCD Error message and the cause of error correspondence table
Table 2 LCD Error message and the cause of error correspondence table
LCD Error message
The cause of error
“E0 1”
Due to leaks too fast or pulse signal is too
weak,please check wristbands,and after tied
please measure again.
“E0 2”
Because there is too much noise,the blood
pressure signal can not be measured,please
remove the noise source,and then measure
again.
“E0 3”
Test results are not normal,please measure
again.
“E0 P
If aerate fails, please check the wristband,
re-tied, and then measured again.
“HI”
When the aerate pressure is higher than 307 ±
8mmHg(41kPa± 1kPa), please measure again.