EasyManuals Logo

Braun No touch + forehead User Manual

Braun No touch + forehead
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
English
11
Care and cleaning
The thermometer should be cleaned in between uses, as necessary. Use
an alcohol swab or cotton swab moistened with alcohol (70% Isopropyl)
to clean the thermometer casing and the measuring probe. Ensure that
no liquid enters the interior of the thermometer. Never use abrasive
cleaning agents, thinners or benzene for cleaning and never soak the
device in water or other cleaning liquids. Wait 10 minutes after cleaning
before taking a temperature measurement. Make sure to replace
the protective cap after use to prevent scratches or damage to
the thermometer.
Never insert a sharp object into the scanner area or any other open
surface on the thermometer.
Clean sensor mirror before each use or as needed.
Replacing the batteries
The Braun No touch + forehead thermometer comes with 2 AA batteries.
Replace with 2 new AA batteries when the flashing battery symbol
appears on the LCD display. To change the batteries, slide open the
battery cover and remove batteries. Replace the batteries being sure
to align properly as indicated inside the battery compartment. Remove
the battery from the product if it is not required for extended periods of
time in order to avoid damage to the thermometer resulting from a
leaking battery.
To protect the environment, dispose of the product and empty
batteries at appropriate collection sites according to national or

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun No touch + forehead and is the answer not in the manual?

Braun No touch + forehead Specifications

General IconGeneral
Measurement MethodInfrared
Backlit DisplayYes
Silent ModeYes
Color-coded DisplayYes
Measurement SiteForehead
Display TypeLCD
MemoryYes
Power SourceBattery
Measurement AreaForehead
Measurement DistanceUp to 5 cm
Measurement Time2 seconds
Accuracy±0.2°C (±0.4°F)
Last Reading MemoryYes
Auto Power OffYes
Fever IndicatorYes
WaterproofNo
Temperature Range34.4 - 42.2 °C (93.9 - 107.9 °F)
Battery2 x AA