EasyManua.ls Logo

Braun No touch + forehead - Errors and Troubleshooting; Low Battery Indicator; Temperature Range Errors; No Power Troubleshooting

Braun No touch + forehead
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Errors and troubleshooting
Situation
When 20% of battery
life is left, the display
flashes the low battery
warning symbol. The
device can still work
until the battery life
has 0% left.
If the steady battery
icon is the only
symbol shown on the
display, the device
cannot work. The
battery should be
replaced immediately.
This message displays
when measured
temperature is lower
than 93.9 °F (34.4 °C)
or higher than 108 °F
(42.2 °C) or when the
room temperature is
outside the operating
range of 59 °F-104 °F
(15°C-40°C)
.
Blank display.
Thermometer does
not have power.
Error Message
Solution
Replace battery.
Replace battery.
Make sure the
protective cap is
removed. Re-measure
the temperature again,
carefully following
the instruction in
“How to use” section.
Please check if the battery
has been loaded correctly.
Also check polarity (<+>
and <–>) of batteries.
Contact Consumer
Relations if thermometer
still does not function.