Example:
Program Mode should only be performed by an experienced user being familiar with
the principles of the Program Mode function and properly trained in using the present
device.
Note: The active Program Mode function always displays this icon in the Display
( ).
Note: Bolus function is disabled for Program Mode.
Set Profile Parameters: The therapy can be started directly via the drug library or via
the Main Menu/Special functions.
Starting Program Mode via Drug Library:
Note: Program Mode settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
Switch on pump with
o and wait until self-check is finished.
Insert disposable and use the drug library according to Instructions for Use.
Select the desired drug with
t and press l.
Select Program Mode with
l.
In the following screen the user has to confirm the number of steps for the
therapy with
k.
The settings for the steps of the infusion are shown on the display. These
settings, configured in the Drug List Editor, need to be confirmed with
r.
To change the values, press
l to change and k to confirm.
Adjust VTBI with
q.
SPECIAL FUNCTIONS
Chapter 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml