TECHN ICAL DATA
Type of unit Infusion Syringe Pump
Classification (acc. to IEC/EN 60601-1) Defibrillator–proof; CF equipment
Protective Class II; Protective Class I in
combination with SpaceStation
Class (acc. to Directive 93/42 EEC) IIb
Moisture protection IP 22 (fluid protected for horizontal usage)
External power supply:
• Rated voltage Via B. Braun SpaceStation or optional
mains adaptor (rated voltage 100 … 240 V
AC~, 50/60 Hz) for stand alone operation
• External low voltage 11 … 16 V DC via Connection Lead SP
12 V or via SpaceStation
Staff call Max. 24 V / 0,5 A / 24 VA (VDE 0834)
EMC IEC/EN 60601-1-2 / 60601-2-24
Time of operation 100 % (continuous operation)
Operating conditions:
• Relative humidity 30 % … 90 % (without condensation)
• Temperature +5 … +40 °C
• Atmospheric pressure 500 … 1060 mbar
Storage conditions:
• Relative humidity 20 % … 90 % (without condensation)
• Temperature -20 … +55 °C
• Atmospheric pressure 500 … 1060 mbar
Type of battery pack (rechargeable) NiMH
Operating time of rechargeable battery Approx. 8 hours at 25 ml/h
Recharging time Approx. 6 hours
Weight Approx. 1.4 kg
Dimensions (W x H x D) 249 x 68 x 152 mm
Volume preselection 0.1 - 99.99 ml in increments of 0.01 ml
100.0 – 999.0 ml in increments 0.1 ml
1000 – 9999 ml in increments 1 ml
Time preselection 00:01 – 99:59 h
Accuracy of set delivery rate ± 2 % according to IEC/EN 60601-2-24
Occlusion alarm pressure 9 levels up to 1.2 bar
Max. Volume in case of single fault For incorrect dosages of 0.1 ml due to
condition malfunctions of the device the pump will
automatically shut off.
Technical inspection (safety check) Every 2 years
TECHNICAL DATA