3.11 Take Over Mode (TOM)
Take Over Mode is a feature to support the user during syringe changes by auto-
matically starting a second Perfusor® Space pump when the first has run empty.
The second pump automatically takes over the infusion rate from the first pump.
Activation:
Start an infusion of the desired medication from the drug library on a
Perfusor® Space pump (see Section 3.3).
Place a second a Perfusor® Space pump in an adjacent slot of the
SpaceStation (either above or below).
Note: Make sure that the pumps are correctly inserted in the SpaceStation.
Navigate to the Drug Library on the second pump (Note: The drug library can
be started over the Start Up or Special Functions Menu).
Navigate through the list with t and select the Care Unit with l. The Care
Unit of the second pump must be the same as the first.
Note: If you have already set the Care Unit once on your pump this step will
be skipped for the next time.
Navigate through the list with t and select the patient profile with l. The
patient profile of the second pump must be the same as the first.
Note: If no profile is set, this step will be skipped.
SPECIAL FUNCTIONS
Chapter 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
BrandBraun
ModelPerfusor Space
CategoryHealth Care Products
LanguageEnglish