If a battery, which is not completely discharged, is charged several times, its
capacity can be reduced.
Under normal temperature conditions a battery can be charged and discharged
approx. 500 times before its lifetime decreases.
When the pump is not connected to mains power the battery discharges itself
slowly. This can occur even when the pump is not operating.
The battery operating time can only be realized if the pump operates
continuously with a fully charged battery at room temperature.
Battery maintenance:
To accurately balance the battery capacity a cyclical battery maintenance is necessary.
The frequency of battery maintenance may be set in the service tool. The battery main-
tenance mode detects a possible capacity loss (e.g. through ageing of the battery pack)
and then the capacity/running time will be calculated anew. After a longer storage
time or a longer operation without battery maintenance it can happen that the battery
pre-alarm time can no longer be maintained. In this case it is necessary to perform a
battery maintenance.
To initiate the discharge process the message „Battery maintenance and the
k-key
will be displayed after switching the pump off. By pressing
k and u the discharge
process will start. The process is interrupted by switching the pump on again. If the
battery maintenance is to be continued a new activation is necessary. After completely
discharging the battery it will be completely charged again. The total duration of the
battery maintenance process takes approx. twelve hours.
Caution: Please take into account that, if the battery maintenance has not been
completed there is a possibility of a reduced battery operating time.
Replacing batteries:
The Battery Pack SP can be exchanged by any user. No special qualification is required.
All rechargeable batteries exhibit a reduction in capacity as they age. This aging is de-
pendant on several factors including charging cycles, temperature and battery usage.
It is recommended to periodically check the function of the battery. A battery should
no longer be used if a change of syringe leads to a Battery nearly empty“ or a “Battery
emptyalarm when it is fully charged.
Caution: Batteries may explode or leak if they are opened or incinerated. Consider
disposal directions!
BATTERY OPERATION AND MAINTENANCE
Chapter 7

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml