77
Check regularly
Check for cleanliness, completeness and damage. Use only according to
Instructions for Use. During an exchange interval of the disposable the pump
has to perform a self-test. Check the following items each time the pump is
switched on: self-check, audible alarm, process- and alarm control indication.
Disinfecting
Caution: Before disinfecting the pump, always disconnect the pump
from the patient, switch off the device and disconnect from power and
other devices (e.g. staff call).
Clean all exposed surfaces using a clean, soft, lint-free cloth dampened with a
mild cleaning solution of warm, soapy water. Make sure to remove any visible
residue from all surfaces prior to disinfecting. Do not spray disinfectants directly
on the pump, use a soft, low lint cloth dampened but not saturated with pro-
duct. After cleaning and disinfecting allow device to dry for at least 20 minutes
prior to use. Wipe magnifying- and displayglas on front of pump door only with
a soft cloth.
Note: Do not use Hexaquart® or other alkylamine containing disinfectants.
Recommended: disinfectant for wiping available from B. Braun: Meliseptol®
Foam pure, Melsitt 10% and Melsept SF 10%.
Note: Keep instrument upright and do not allow any part of instrument to
become saturated with or submersed in fluid during cleaning operation.
Do not allow moisture or detergents to come into contact with the electrical
connections of the device (P2 or P3 connectors) or any device openings. To
reduce the likelihood of moisture ingress into the electrical connectors, the P2
connector of a power supply or combi cable may be used to cover the
connections during cleaning operations. Ensure that any connectors used to
cover are not connected to a wall outlet or other electrical source. Once the
cleaning has been completed, remove the connector and inspect all connectors
for residual moisture and evidence of damage or breakdown to the plating on
the connectors. Allow any residual moisture to evaporate before plugging the
device into a wall outlet. Replace any connectors which exhibit damage or
evidence of plating breakdown prior to returning the device to service.
Utilize electrical contact cleaner that does not react with plastics to remove any
deposits of material which may be present inside the electrical connectors as
required.
WARRANTY / TSC* / SERVICE / TRAINING / DESINFECTING / DISPOSAL
Chapter 11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml