83
Chapter 12
INSTRUCTIONS FOR USE ACCESSORY
PCA-Accessories
• Space PCA-Kit (REF 8713554) consisting of: :
- Demand button
- Hook and loop tape
f
or fixation of the demand button at
the patients arm
- Line fixation
connection between hook and loop tape
and demand button
- Metal clip
alternatively for fixation at the bed
sheet
- Cable strap
for wrapping the cable of the demand
button
- PCA-Key
for locking the syringe holder or the Syringe Anti Removal Cap
• Syringe Anti Removal Cap PSP (REF 8713556)
Fixation of the demand button:
at the wrist: or at the bed sheet:
Usage of the cable strap:
Usage of the Syringe Anti Removal
Cap PSP:
The Syringe Anti Removal Cap PSP is
slided over the drive head from the
front and is fixed with the PCA-key
(270° clockwise rotation). Mind the
markings – make sure it is securely locked. Dismantling: counter clockwise
rotation of 270°. Push to the left and disengage.
Caution: When Syringe Anti Removal Cap is used always change the syringe as
soon the „syringe empty“ alarm appears.