11
English
Errors and troubleshooting
Error message Situation Solution
When 20% of battery life is left, the
display ashes the low battery warning
symbol; however, the device can still
work until the battery life has 0% left.
Replace batteries.
If the steady battery icon is the only
symbol shown on the display, the device
cannot work. The battery should be
replaced immediately.
Replace batteries.
This message displays when the mea-
sured temperature is lower
than 34 °C (93.2 °F) or higher than
43 °C (109.4 °F) or when the room tem-
perature is outside the operating range
of 15 °C - 40 °C (59 °F - 104 °F).
Re-measure the temperature,
carefully following the instruc-
tions in “How to use” section.
System error – self-check display ashes
continuously and will not be followed by
the ready beep and the ready symbol.
If error persists,
If error still persists,
Wait 1 minute until the thermom-
eter turns o automatically, then
tur
n on again.
... reset the thermometer by
removing the batteries and
putting them back in.
… please contact Consumer
Relations.
Blank display. Thermometer does not
have power.
Please check if the batteries
have been loaded correctly.
Also check polarity (<+> and
<–>) of batteries.
Contact Consumer Relations
if thermometer still does not
function.
English