9 Maintenance Instructions
NIPPY Clearway 2 Service Manual
2210.2657 Rev 3.0
Inspection Equipment and Tools
The following equipment is required to perform an inspection on the
Clearway 2:
Gas flow analyser.
or
Manometer capable of measuring at least ±70 cmH
2
O (± 68.6 hPa) (±
0.75%) and a flow meter capable of measuring at least 300 sl/m (± 2%).
70% Alcohol cleaning wipes
Test Lung (part no 6011)
6mm leak port (part no 00997)
Clearway Breathing Circuit (part no 0960.101)
Medical Electrical Safety tester
The following tools are required to open the case, remove and replace
components when necessary:
Torx screwdriver T15 x 200mm (To open the case)
Pozidrive screwdriver No. 1
Hex screwdriver 2mm
Hex screwdriver 2.5mm (To open the battery cover)
Soldering iron, to remove and replace the alarm battery every 5 years