22 23
Technical Specifications
30W Power Adapter Unit Input range 100–240V, 50–60Hz
Typical power consumption 20W (30VA)
Maximum power consumption 35W (70VA)
Operating temperature 41
º
F to 95
º
F (5
º
C to 35
º
C)
Operating humidity 10–80% Relative Humidity, Non-Condensing
Operating altitude Sea level to 8,000 ft (2,433 m)
Operating atmospheric pressure 101 to 75.3 kPA
Storage and transport temperature -4
º
F to +140
º
F (-20
º
C to +60
º
C)
Storage and transport humidity 10–95% Relative Humidity, Non-Condensing
Storage and transport altitude No limit
Storage and transport atm. pressure No limit
Electromagnetic compatibility Compatible with IEC 60601-1-2
Aircraft use Complies with RTCA DO-160G Standard for use in
flight
IEC 60601-1 classification Class II (double insulation), Type BF
Available modes Standby, CPAP and CPAP with ramp
Deliverable pressure accuracies ±0.6 cm H
2
O ± 4% of the measured reading
Sound pressure level 26 dBA
Nominal dimensions (L x W x H) 6.48” X 3.30” X 2.02” (16.46 cm x 8.38 cm x 5.13 cm)
Weight 10.3 oz. (292 g)
Housing construction Flame retardant engineered thermoplastic
Supplemental oxygen Not for use with oxygen
Standard air filter Polyester
Air outlet The 22 mm air outlet complies with ISO 5356-1
Battery charges when Z2 not in use.
Performance Specifications
Operating pressure range 4 to 20 cm H2O
Maximum single fault steady pressure 30 cm H2O
Pressure accuracy ±(0.6 cm H2O + 4% of the measured reading)
Bluetooth Specifications
Panasonic Electronic Components
Bluetooth Module + BLE HCI W/ANTENNA
FCC ID T7V1315
FCC Compliance FCC PART 15 SUBPART C
Frequency 2400-2483.5 MHz
Modulation or Protocol Bluetooth Low Energy, V4.0
Over the air data rate 1 Mbit/s
Application throughput 0.27 Mbit/s
Security 128-bit AES with Counter Mode CBC-MAC and
application layer user defined
Robustness Adaptive frequency hopping, Lazy Acknowledge-
ment, 24-bit CRC, 32-bit Message Integrity Check
Latency (from a non-connected state) 6ms
Power-Output 10dBm
Sensitivity -93dBm
Standard IEEE 802.15.1
The following symbols may appear on your device, power supply unit, or packaging.
Refer to Manual type BF equipment; Class II equipment; Start/stop;
Ingress Protection (IP) Rating; Manufacturer; Date of manufacture; Batch
code; Catalog number; Serial number; Direct current; Disengagement
Mechanism; Prescription only (In US. Federal law restricts these device to sale by or
on order of a physician); FCC ID FCC Identifier.