5 SPECIFICATIONS
Capacity: 500g x 0.01g, 3kg x 0.1g,
Units: g/oz/lb/kg/tl/ct
Tare range: Tare full capacity
Auto off (default): 60 seconds off
Operation temperature: 5-35 degrees
Display: LCD 5 digits
6 CALIBRATION
Test weights needed: 500g cap. = 200g & 500g, 3kg = 1kg & 3kg
Press to turn the scale on.
Press and hold the M key until CAL is displayed. Display briefly shows CAL and then flashes a weight (smaller test weight)
to be placed on the scale.
Add all the requested weight at the same time to the scale. The display will then show a second weight equal to capacity of
the scale.
Remove first weight and place the larger weight all at once to the scale. The display shows PASS, then the total weight.
Remove the test weights and press to turn off scale. The calibration is complete.