MS140/300
Portable Medical Electronic Handrail Scale
User Instructions
AWT35-501200 Issue AA June 2013

Table of Contents

Questions and Answers

  • S
    Stephanie WeeksAug 14, 2025
    What does 'CAL.Er' mean on my Brecknell MS140/300 Scales and how to fix it?
    • C
      Christopher RomeroAug 14, 2025
      'CAL.Er' on your Brecknell Scales indicates a calibration error. This can occur if the input data or loaded weight is too small, too big, unstable, or non-linear. Input the correct data and load the correct weight onto the platform. If the issue persists, service is required.
  • A
    Ashlee JonesAug 18, 2025
    Why won't my Brecknell MS140/300 turn on?
    • J
      Janet FullerAug 18, 2025
      The Brecknell Scales may not turn on due to several reasons: the power cord might not be properly connected or plugged in, the power outlet may not be supplying electricity, or the battery might be discharged. To resolve this, check the power cord connections and ensure it's plugged into a working power outlet. Verify the power source and replace the batteries if necessary. If the issue persists, service is required.
  • J
    Jerry TaylorAug 21, 2025
    What to do if my Brecknell MS140/300 display 'Ad------'?
    • E
      ejosephAug 21, 2025
      If your Brecknell Scales display 'Ad------', it indicates that the loadcell wires to the indicator are incorrectly connected, shorted, or opened, or that the ADC or loadcell(s) are damaged. Ensure the wires are correctly connected and in good condition. If the problem persists, the loadcell or ADC chip may need replacement, which requires professional service.
  • K
    Kelly RodriguezAug 25, 2025
    What to do if my Brecknell MS140/300 Scales cannot zero the display?
    • C
      Christian SimmonsAug 25, 2025
      If your Brecknell Scales cannot zero the display or will not zero when turned on, it could be due to several reasons. The load on the scale might exceed allowable limits, the load on the scale might not be stable, or there might be load cell damage. Remove any load on the scale and wait for the load to become stable. If the issue continues, service is required.
  • T
    Tanner JoyceAug 27, 2025
    What does 'EEP.E2' mean on my Brecknell MS140/300 and how to resolve it?
    • C
      Carl DanielsAug 27, 2025
      The error message 'EEP.E2' on your Brecknell Scales means that a parameter is not correctly set. To fix this, re-set the parameter.
  • M
    Matthew RogersAug 31, 2025
    What does '--------' mean on Brecknell MS140/300 Scales?
    • A
      Angela KeyAug 31, 2025
      The '________' message on Brecknell Scales means the weight reading is below the under load limit. Install the platform on the scale and perform a zero calibration.
  • J
    Jennifer HarrisSep 3, 2025
    What does '________' mean on Brecknell Scales?
    • J
      Javier CarterSep 4, 2025
      The '________' message on Brecknell Scales means the weight reading is below the under load limit. Install the platform on the scale and perform a zero calibration.
  • A
    Alexis GouldSep 7, 2025
    What does '0------' mean on Brecknell MS140/300?
    • B
      barbaraclarkSep 7, 2025
      The '0______' message on Brecknell Scales means that the weight reading is below the Power On Zero limit. Install the platform on the scale and perform a zero calibration.
  • A
    Alicia PrattSep 12, 2025
    What does '0______' mean on Brecknell Scales?
    • J
      John KingSep 12, 2025
      The '0______' message on Brecknell Scales means that the weight reading is below the Power On Zero limit. Install the platform on the scale and perform a zero calibration.

Summary

Safety and Maintenance

Operating Warnings

Lists essential operating precautions and safety warnings for the scale.

Scale Specifications

Indicator Specifications Overview

Outlines technical specs for power, dimensions, display, and keypad.

Introduction to Operation

Keyboard Functions in Normal Mode

Details the functions of the four keys on the scale's keypad in normal mode.

Weighing Mode Navigation

Explains operation, powering, HOLD, zero, tare, and clear tare functions.

Setup Mode Configuration

Setup Menu Parameters

Detailed descriptions of various setup parameters available in the menu.

Setup Menu Parameters (Continued)

Details setup parameters P15 through P19, covering data filter, stability, and overload limits.

Calibration Procedures

Calibration Mode Steps

Steps to enter and perform calibration using standard weights.

Calibration Mode Steps (Continued)

Continues detailed steps for calibration, including setting calibration points.

Serial Communication

Commands and Responses Guide

Lists commands for controlling the scale and their expected responses.

Commands and Responses Guide (Continued)

Continues the list of commands and responses for scale communication.

Troubleshooting Guide

Error Messages and Solutions

Lists common error symptoms, probable causes, and recommended remedies.

Error Messages and Solutions (Continued)

Provides more detailed troubleshooting steps for common issues like power and zeroing.