Operation
Temperature .............................................50°F to 104°F
Humidity.......................................................30% to 75%
Electrical Requirements
... .........120 VAC nominal, 60 Hz, 3 amps maximum
... .........including maximum 2 amp accessory outlet
Fuse rating ..............................3 Amp/250V P/N 102007
Power Cord ........ Extends 78 in. (Minimum) from table 3
conductor, 120VAC. With hospital grade plug
Type of protection against electrical shock ..... Class 1
Type of protection
against electrical shock ......................... B applied parts
Type of protection against ingress of water……. Ordinary
Can accept paper rolls of.................. 21 in. x 3.5 in. dia.
Certications*......
UL60601-1;CAN-CSA C22.2 No. 601.1
Pelvic Tilt ..............................................................5° Rise
Drawer Warmer .................120 VAC, 65 watts, 0.5 amps
Weight of Table ...................................................220 lbs.
• With carton and skid ...........................................265 lbs.
Table Weight Capacity .......................................500 lbs.
Stirrup Extension (maximum) .............................. 16 in.
Stirrup Lateral Positions ...................4 Lateral Positions
Leg Support Extension (maximum) ..................... 15 in.
Dimensions
• Upholstered Top ..................28.0 in. wide x 57.3 in. long
• With Leg Support Fully Extended ....................... 71.0 in.
• Seat Height ........32 in to top of upholstered seat at foot
• Step ................ 7.2 in. high x 20.38 in. wide x 11 in. long
Step Capacity ......................................................500 lbs.
Base ............ 28.2 in. high x 23.25 in. wide x 46.0 in. long
Transportation and Storage
Temperature .............................................18°F to 176°F
Humidity......................................................-10% to 90%
Explosion/Fire Hazard
Do not use this table in an explosive or oxygen
rich atmosphere. Failure to do so may result in
serious personal injury or death.
WARNING
Shock Hazard
When performing cauterization or similar
treatment, the patient must be insulated
from the metal portions of the table by non-
conductive material. Failure to do so may
result in electrical shock.
WARNING
Certications
This product has been evaluated with respect to electrical
shock, re, and mechanical hazards only in accordance with
UL60601-1; CAN/CSA C22.2 No. 601.1
NOTICE
Electromagnetic Interference (EMI)
This product is designed and built to minimize electromagnetic
interference with other devices; however, if interference is
noticed between another device and this product, remove the
interfering device from the room or plug this product into an
isolated circuit.
NOTICE
*Classied in the United States and Canada per
the standards listed above.