57
Chapter 7
Specifications
7.1 Compliance with Standards 4128-C and 4128-D
,
CE-mark indicates compliance with: EMC Directive and Low Voltage Directive.
C-Tick mark indicates compliance with the EMC requirements of Australia and New
Zealand
Safety
EN/IEC61010–1: Safety requirements for electrical equipment for measurement, control
and laboratory use.
ANSI/UL61010–1: Safety requirements for electrical equipment for measurement,
control and laboratory use.
EMC Emission
EN/IEC61000–6–3: Generic emission standard for residential, commercial and light
industrial environments.
CISPR22: Radio disturbance characteristics of information technology equipment. Class
B Limits.
FCC Rules, Part 15: Complies with the limits for a Class B digital device.
EMC Immunity
EN/IEC61000–6–1: Generic standards – Immunity for residential, commercial and light
industrial environments.
EN/IEC61326: Electrical equipment for measurement, control and laboratory use – EMC
requirements.
Note: The above is only guaranteed using accessories listed in the Product Data
(BP0512).
Temperature
IEC60068–2–1 and IEC 60068–2–2: Environmental Testing. Cold and Dry Heat.
Operating Temperature: –5C to +40C (+23F to +104F)
Storage Temperature: –25C to +70C (–13F to +158F)
Humidity IEC60068–2–78: Damp Heat: 90% RH (non-condensing at 40 C (104 F))
Mechanical
Non-operating:
IEC60068–2–6: Vibration: 0.3mm, 20m/s
2
, 10–500 Hz
IEC60068–2–27: Shock: 1000 m/s
2
IEC60068–2–29: Bump: 1000 bumps at 250m/s
2