Chapter 7 — Low-frequency Pressure-field
1
/
2
" Microphone Type 4193
Introduction
Falcon™ Range of Microphone Products
Microphone Handbook
BE1379– 12
7
−
3
flush-mounted or inserted into closed volumes as it can be supported by the dia-
phragm clamping ring, provided that a force of less than 5Newtons is applied.
The microphone is supplied with individual calibration data on a calibration chart
and on a 3
1
/
2
" data disk in a case. This case can also contain a
1
/
2
" Microphone
Preamplifier Type 2669.
7.1.2 The Calibration Chart
Each microphone is supplied with an individual calibration chart (see Fig.7.2)
which gives the microphone’s open-circuit sensitivity, polarized capacitance, cut-off
frequency and pressure-field and random-incidence frequency responses. The data is
valid for a microphone without Adaptor UC0211 fitted.
Open-circuit Sensitivity
The stated open-circuit sensitivity is valid at the reference frequency (251.2Hz
*
) for
free-field, random-incidence and pressure-field conditions. The stated uncertainty is
the U
95
value (the value valid for 95% confidence level).
Ambient Conditions
The ambient conditions are measured continuously during calibration at the factory.
The calibration results obtained at the measured Environmental Calibration Condi-
tions are corrected to the reference ambient conditions stated under Valid At (23°C,
101.325kPa and 50%RH).
*The exact reference frequency is 10
2.4
Hz (re ISO 266).
Fig.7.2 Microphone calibration chart
Frequency Hz
dB
+ 5
0
– 5
– 10
– 15
1 2 5 10 20 50 100 200 500 1 k 2 k 5 k 10 k
– 20
20 k 40 k
+ 1
– 1
940954/1e
Dotted Curve Shows Typical Response
Random-incidence Response
See the microphone handbook for further information
Pressure-field Response
BC 0228 – 12
Cut-off Frequency (– 3 dB)
22
mHz
Bruel & Kjær
B
K
7/6-'89
Low-frequency Pressure-field
1/2" Microphone Type 4193
Calibration Chart
Serial No:
4193000
Open-circuit Sensitivity*, S
0
:
-38.0
dB re 1V/Pa
Equivalent to:
12.6
mV/Pa
Uncertainty, 95 % confidence level 0.2
dB
Capacitance:
18.4
pF
Valid At:
Temperature:
23
°C
Ambient Static Pressure: 101.3
kPa
Relative Humidity: 50 %
Frequency: 250
Hz
Polarization Voltage, external: 200 V
Sensitivity Traceable To:
DPLA: Danish Primary Laboratory of Acoustics
NIST:
National Institute of Standards and Technology, USA
IEC 1094-4: Type WS 2 P
Environmental Calibration Conditions:
100.1 kPa 25 °C 35 % RH
Procedure: 704219 Date: 21. Apr. 1994 Signature:
*K
0
= – 26 – S
0
Example: K
0
= – 26 – (– 38) = + 12 dB