200.0V and 600V and then. The AC voltage ranges are
2.000V, 20.00V, 200.0V and 600V.
To measure DC or AC voltage:
Insert the red test lead into the VΩ· input terminal
and the black test lead into the COM terminal.
Set the rotary switch to DC or AC range.
Connect the test lead across with the object under
testing.
The measured value will be show on the LCD
display.
Note:
When DC or AC voltage measurement has been completed,
disconnect the connection between the testing lead and the
circuit under testing.
5-2. Resistance measurement
The resistance range are: 200.0έ , 2.000Kέ ,20.00Kέ ,
200.0Kέ , 2.000Mέ .20.00Mέ .
To measure resistance, connect the meter as follows:
Insert the red test lead into the · · terminal and
the black test lead into the COM terminal.
Set the rotary switch to proper resistance range..
Connect the test lead across with the object under
testing.
The measured value will be show on the LCD display.
Note:
The test lead can add 0.1 έ to 0.2 έ of error to
resistance measurement. To obtain precision reading
in low-resistance measurement, that is the range of
200.0έ , short the input terminal before measuring.
In this time, the contact resistance displayed on the
LCD. You can subtract the contact resistance value
from the measured value.