3 Technical data
11 F-100 / F-105 / F-108 / F-114 Operation Manual, Version E
3 Technical data
This section introduces the reader to the Recirculating Chiller and its specifications. It contains the
technical data, requirements and performance data.
3.1 Technical data
Technical data (230 V)
F-100 F-105 F-108 F-114
Power consumption (max.) 850 W 850 W 1350 W 1850 W
Supply voltage 230 VAC ±10% 230 VAC ±10% 230 VAC ±10% 230 VAC ±10%
Fuse 6.3 AT 6.3 AT 7 AT 8 AT
Frequency 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz
Environmental conditions For indoor use only
- Temperature 5–35°C 5–35°C 5–40°C 5–40°C
- Altitude up to 2000 m
- Humidity Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to
50% relative humidity at 40°C
Display No display digital, resolution
0.1°C
digital, resolution
0.1°C
digital, resolution
0.1°C
Overvoltage category II II II II
Degree of protection IP20 IP20 IP20 IP20
Pollution degree 2 2 2 2
Instrument dimensions
W×H×D [mm]
280×500×420 280×500×420 400×580×500 400×660×500
Weight 28 kg 30 kg 40 kg 42 kg
Cooling capacity at 15°C --- 530 W 800 W 1400 W
Cooling capacity at 10°C 300 W 390 W 650 W 1100 W
Cooling capacity at 0°C --- 120 W 400 W 700 W
Cooling capacity at -10°C --- 10 W 150 W 350 W
Refrigerant R134 R134 R134 R134
Temperature range x +10°C –10°C … +25°C –10°C … +25°C –10°C … +25°C
Accuracy ±2°C ±1°C ±1°C ±1°C
Tank volume 3 L 3 L 4.5 L 6.6 L
Hose connection 8 mm 8 mm 9.5 mm 13.5 mm
Pump ow rate 2.5 L/min 2.5 L/min 3.0 L/min 11.0 L/min
Pump pressure 0.6 bar 0.6 bar 0.6 bar 1.0 bar