4 Controls and connections
4.1 Controls (Mounting rail version)
4.1.1 Front panel
The front panel contains two bores for accessing the potentiometers for fine adjustment of the
zeropoint and the gain.
4.1.2 Rear panel
The rear panel contains two bores for accessing the potentiometers for fine adjustment of the
sensor feed voltage and the 5-mV calibration source. The sensor feed voltage only requires
calibration by the customer after it has been switched from the factory default setting (5 V) to a
different value. In this case, the feed voltage can deviate by up to 0.2% from the setpoint value.