77
Safety Designation per IEC 60601-1
Type of protection against
electrical shock
Class II
Degree of protection against
electrical shock
Type B applied part
Supply connection 100-240 VAC, 50/60Hz
Power Equipment Adapter and rechargeable
internal NiMH battery.
Battery life: Approximately 30 hours with
a fully charged new battery.
Degree of Electrical connection
between equipment and Patient
Equipment designed as non-
electrical connection to the
patient.
Degree of mobility Transportable
Mode of operation Continuous
NOTE: When you connect other equipment to the unit, always make
sure that the whole combination complies with the international safety
standard IEC 60601-1 for medical electrical systems. When
connecting to a PC with the supplied USB lead then the PC must be
EN 60601-1 / ANSI/AAMI ES60601-1:2005 compliant.
IMPORTANT: Only use the mains adapter supplied (36-PSU1017 5V
DC 1.2A). The adapter contains a transformer. Do not cut off the
adapter to replace it with another plug as this causes a hazardous
situation.
• The adapter transforms the mains voltage (100-240 Volts) to a safe
voltage (5V DC)
• Make sure the adapter does not get wet
• Do not use a damaged adapter
• Always unplug your Micro I before cleaning
WARNING: Do not connect devices that are not specified as part of
the system.
WARNING: No modification of this equipment is allowed.