Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Micro I is intended for use in the electromagnetic environment specified
below. The customer or the user of the Micro I should assure that it is used in
such an environment.
Immunity
Test
IEC 60601 test
level
Compliance
Level
Electromagnetic
Environment –
Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2 kV for
power supply
lines
+/- 1 kV for
input/output
lines
+/- 2 kV for
power supply
lines
[Not Applicable]
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV line(s)
to line (s)
+/- 2 kV line(s)
to earth
+/- 1 kV line(s)
to line (s)
[Not Applicable]
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
< 5% UT
(>95% dip in UT)
for 0.5% cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5s
< 5% UT
(>95% dip in UT)
for 0.5% cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5s
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the Micro
I requires continued
operation during
power main
interruptions, it is
recommended that the
Micro I be powered
from an
uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristics of a
typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.