125
DIAGRAMMA DI LAVORO DI Micro-Smart
Micro Smart WORKING DIAGRAM
Fig. F
Draw.F
70
4
(66)
26
(433)
250
210
ASPIRAZIONE mbar
ASPIRATION mbar
100
0
50
150
0 10
(166)
20
(333)
30
(500)
PORTATA m3/h
(l/min)
FLOW RATE m3/h (l/min
)
CURVA ASPIRAZIONE MICRO-SMART (SET POINT VUOTO 70 mbar)
PREVALENZA CONSIGLIATA PER ASPIRASALIVA o ASPIRAZIONE DI FONDO BOCCA.
ASSORBIMENTO IN (A) Da 0,5 a 1 - INTERVALLO DI FREQUENZA (Hz) Da 50 a 55
MICRO SMART ASPIRATION CURVE (VACUUM SET-POINT 70 mbar)
SUGGESTED MAX.HEAD FOR SALIVA EJECTOR OR MOUTH FLOOR ASPIRATION
ABSORPTION IN (A) from 0.5 to 1 –CYCLE RANGE (Hz) from 50 to 55
CURVA ASPIRAZIONE MICRO-SMART (SET POINT VUOTO 100 mbar)
PREVALENZA CONSIGLIATA PER INTERCETTATORE DI SPRUZZI.
ASSORBIMENTO IN (A) Da 1,1 a 1,9 - INTERVALLO DI FREQUENZA (Hz) Da 60 a 90
MICRO SMART ASPIRATION CURVE (VACUUM SET-POINT 100 mbar)
SUGGESTED MAX.HEAD FOR SPRAY INTERCEPTOR
ABSORPTION IN (A) from 1.1 to 1.9 –CYCLE RANGE (Hz) from 60 to 90
CURVA ASPIRAZIONE MICRO-SMART (SET POINT VUOTO 210 mbar)
PREVALENZA CONSIGLIATA PER ASPIRAZIONE CHIRURGICA.
ASSORBIMENTO IN (A) Da 3,1 a 3,6 - INTERVALLO DI FREQUENZA (Hz) Da 95 a 105
MICRO SMART ASPIRATION CURVE (VACUUM SET-POINT 210 mbar)
SUGGESTED MAX.HEAD FOR SURGICAL ASPIRATION
ABSORPTION IN (A) from 3.1 to 3.6 –CYCLE RANGE (Hz) from 95 to 105
Escludendo i tasti preselezionati, Micro-Smart può lavorare
su due poltrone contemporaneamente.
Excluding the preset buttons, Micro-Smart can operate two
chairs at the same time.