EasyManua.ls Logo

CEFLA NewTom 5G - Page 63

CEFLA NewTom 5G
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Quality assurance
as shown in the following picture.
4) Finely adjust the phantom position by using the table console buttons. (UP/DOWN LEFT/RIGHT
[FORWARD/BACK – patient table only] ).
In order to help the operator during the positioning of the patient the laser can be turned on by
pressing the LASER button on the scanner control panels or on the table console (required the NNT
software to be open).
Position the phantom in such a way the laser will match with the reference line located on the
phantom (see next picture=.
5) For the phantom scan please refer to Par. “QA Phantom scan” and Remote patient position
adjustment” included in the “Acquisition Operations with NewTom VGi-5G” document annex to the
“NNT User Manual” document.
7-2 NewTom 5G – User Manual

Table of Contents