BIOPRINTING WITH CELLINK SOFTWARE (REPETIER
HOST)
INTRODUCTION TO G-CODE PROGRAMMING
The special M codes for INKREDIBLE 3D Bioprinter
M712 to M732 are used to set the offset between the printheads
printhead 1 and 2 are the syringes and printhead 3 is the led.
Note: That this will not place the chosen printhead in the active position.
M712 Set offset from 1 to 2
M713 Set offset from 1 to 3
M721 Set offset from 2 to 1
M723 Set offset from 2 to 3
M731 Set offset from 3 to 1
M732 Set offset from 3 to 2
M750 homes the printheads. This needs to be done before any automatic switching of the
printheads can occur. After this command Printhead 1 will be in the active position
M751 Places Printhead 1 in the active position. This is faster than M750.
M752 Places Printhead 2 in the active position
M753 retracts both Printhead 1 and 2 making way for the led to be used.
M760 Opens the valve for printhead 1
M761 Closes the valve for printhead 1
M762 Opens the valve for printhead 2
M763 Closes the valve for printhead 2
M764 Turns on the UV LED
M765 Turns off the UV LED
Workflow
Import the models into the slicing software and slice.
Open the G-code file and make sure to remove any homing command in the start of the code.
INKREDIBLE needs to be homed manually from the menu to account for different print heights.
NOTE: FOLLOW THE CALIBRATE PRINTBED SEQUENCE ABOVE.