OPERATING THE INKREDIBLE
Only trained staff is allowed to operate the instrument being aware of the risks and potential
injuries (non reversible damages can be caused).
NOTE: OPENING OF THE INSTRUMENTS ELECTRICAL PANEL WILL AUTOMATICALLY
REVOKE THE WARRANTY.
The manufacturer certifies that the INKREDIBLE is conforming to requirements.
The manufacturer gives an 8 months warranty on the moving parts in the INKREDIBLE
bioprinter. The guarantee excludes deliberate damages.
SETUP AND INSTALLATION
1. Unpacking the 3D Bioprinter
Once the INKREDIBLE 3D Bioprinter has been carefully taken out of the box, all protective films
plastic parts preventing the axis to move during transportation can be removed. Once all
protective materials and films have been removed, the INKREDIBLE can be placed inside a
biological hood or laboratory bench. The INKREDIBLE should never be placed on
wet/damp/moist/humid surfaces. Make sure to remove all BLACK zip ties holding the X-
carriage and Printbed in place during transportation.
2. Remove the Accessory Box and the Air Supply Box
3. Place the INKREDIBLE bioprinter inside the biological hood
4. Calibrating the printbed for the first time (every time you move the INKREDIBLE you
should follow this calibration procedure):
A standard leveler should be used for calibrating the printbed for the first time.
Place the leveler at the back of the printbed and adjust the back screws to level.
Place the leveler on the middle of the printbed and perpendicular to the back screws. Adjust the
front screw to level.
5. The INKREDIBLE 3D Bioprinter can be placed inside an isolator
6. The power cord can be plugged in the back [9] of the of the INKREDIBLE 3D bioprinter
and then in the outlet inside of the isolator or outside the isolator. Note: Ensure to always
plug in the power cord into a separate individual outlet for bioprinter.