EasyManua.ls Logo

CEM DT-610B - User Manual

CEM DT-610B
2 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
¡ ã
Tc
C
¡ã
K
F
¡ ã
BACK-
LIGHT
10. ¡ ã
+
MAX HOLD
F¡ã
K
C¡ ã
OPERATING INSTRUCTIONS
DIGITAL THERMOMETER
INTRODUCTION
This instrument is a portable 3
1
/
2
digit., compact-sized
digital thermometer designed to use external K-type
thermocouple as temperature sensor. Temperature
indication follows National Bureau of Standards and
IEC584 temperature/voltage tables for K-type
thermocouples. One K-type thermocouples is supplied
with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and operation
instructions before using the thermometer.
WARNING
TO AVOID ELECTRICAL SHOCK. DO NTO USE THIS
INSTRUMENT WHEN VOLTAGES AT THE
MEASUREMENT SURFACE EXCEED 24V AC OR 60V
DC.
WARNING
TO AVOID DAMAGE OR BURNS. DO NOT MAKE
TEMPERATURE MEASUREMENTS IN MICROWAVE
OVENS.
CAUTION
Repeated sharp flexing can break the thermocouple leads.
To prolong lead life, avoid sharp bends in the leads,
especially near the connector.
The symbol on the instrument indicates that the
operator must refer to “Input protection” in this manual.
SPECIFICATIONS ELECTRICAL
Temperature Scale:
Celsius ( ), Fahrenheit ( ) or Kelvin (K)
user-selectble
Measurement Range:
-50 to 1300, -58 to 2000, 223K to 2000K
Resolution: 1 or 1, 0.1 or 0.1,1 K
Accuracy:
Accuracy is specified for operating temperatures over the
range of 18to 28(64 to 82), for l year, not including
thermocouple error.
±2---------------------- -50 to 0
±4---------------------- -58 to 32
±5K---------------------- 223K to 273K
±(0.5% rdg+1)-------- 0 to 1000
±(0.8% rdg+1)-------- 1000 to 1300
±(0.5% rdg+2)-------- 32to 2000
±(1.0% rdg+2K)-------- 273Kto 2000K
Temperature Coefficient:
0.1 times the applicable accuracy specification per
from 0 to 18 and 28 to 50 (32 to 64 and
82 to 122).
Input protection:
60V dc or 24V rms ac maximum input voltage on any
combination of input pins.
Reading Rate: 2.5 times per second.
Input Connector:
Accepts standard miniature thermocouple connectors (flat blades
spaced 7.9mm, center to center).
ENVIRONMENTAL
Ambient Operating Range:
0 to 50 (32 to 122)
Storage Temperature:
-20 to 60 (-4 to 140)
Relative Humidity:
0% to 80% (0 to 35) (32 to 95)
0% to 70% (35 to 50) (95 to 122)
GENERAL
Display:
3
1
/
2
digit liquid crystal display (LCD) with maximum
reading of 1999
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22)
Dimensions:
162mm (H)×76mm(W)×38.5mm(D)
Weight: 210g
Supplied Probe:
4 foot type"K"thermocouple bead probe (Teflon tape
insulated).
Maximum insulation temperature 260(500)
Probe accuracy ± 2.2 or ± 0.75% of reading
(Whichever is greater) from 0°to 800
OPERATING INSTRUCTIONS
Selecting the Temperature Scale
Readings are displayed in either degrees Celsius (),
degrees Fahrenheit () or degrees Kelvin (K).When the
thermometer is turned on, it is set to the temperature scale
that was in use when the thermometer was last turned off.
To change the temperature scale, press the For,or K
key.
Question and Answer IconNeed help?

Do you have a question about the CEM DT-610B and is the answer not in the manual?

CEM DT-610B Specifications

General IconGeneral
Measurement TypeTemperature and Humidity
DisplayLCD
Power Supply9V battery
Resolution Humidity0.1% RH
Accuracy Temperature±1.0°C
Resolution Temperature0.1°C
Temperature Range-10°C to 50°C (14°F to 122°F)

Summary

Introduction to Digital Thermometer

Safety Information and Warnings

Electrical Specifications

Temperature Scale Options

Selectable temperature scales: Celsius (°C), Fahrenheit (°F), or Kelvin (K).

Measurement Range and Resolution

Specifies the measurement range and available resolutions for the thermometer.

Accuracy Specifications

Details accuracy under specified operating temperatures and for one year.

Input Protection and Reading Rate

Defines maximum input voltage and the rate at which readings are taken.

Input Connector Type

Describes the standard miniature thermocouple connector.

Environmental Operating Conditions

Ambient Operating and Storage Range

Specifies the acceptable temperature and humidity for operation and storage.

General Instrument Details

Display, Battery, and Physicals

Information on the LCD display, battery type, dimensions, weight, and probe.

Operating Instructions and Modes

Selecting Temperature Scale and Resolution

Guidance on how to select the desired temperature scale and display resolution.

Overload Display and Hold Mode

Explains overload indication and how to use the data hold function.

Max Mode and Backlight

Instructions for using maximum value recording and the LCD backlight.

Temperature Compensator Checking

How to enter the mode to check internal temperature compensation.

Operator Maintenance

Battery Replacement Procedure

Step-by-step instructions for replacing the 9V battery.

Related product manuals