Non-Contact Forehead InfraRed Thermometer
Table 204-Declaration-Electromagnetic Immunity-
IEC 60501
Test Level
3 Vrms
150k to 80MHz
3 V/m
80M to 2.5GHz
Guidance and-Manufacturer’s Declaration. Electromagnetic
Immunity
The DT-8807H is intended for use in the electromagnetic environment
specified below.
The customer or the user of the DT-8807H should assure that it is
used in such an environment.
Immunity
Test
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
Compliance
Level
3 Vrms
3 V/m
Electromagnetic
Environment-Guidance
Portable and mobile RF
communications
equipment should be used
no closer to any part of
the DT-8807H paration
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
d=1.2 P
d=1.2 P ,
80MHz to 800MHz
d=2.3 P ,
800MHz to 2.5MHz
Where P is the maximum output
power rating of the transmitter
In watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
a
an electromagnetic site survey,
should be less than the
compliance level in each
b
frequency range.
Interference may occur In the
vicinity of equipment marked
with the following symbol: