Clamp Meter
2
Power Source:
1.5V size AAA battery X 2
Typical battery Life: (without buzzer, backlight, flashlight function)
Type:
30 hours at ACA function;
60 hours at ACV function;
100 hours at DCV and Ohm function.
Auto Power Off:
If there is no key or dial operation for 30 minutes, the meter will power
itself off to save battery consumption. This function can be disabled by
press and hold the “ HOLD ” button then power the unit on
Over Load:
When the signal larger than the maximum will be show
.
Maximum jaw opening:
13 mm
Dimensions:
220 x 72 x 35 mm
Weight:
210g (with battery)
Accessories:
Carrying case, Batteries, Test Lead & Instruction Manual.
3. ELECTRICAL SPECIFICATION
The accuracy specification is defined as ±(…%reading…count ) At
23±5°C, 80 %RH.
3-1 Direct Voltage
Range Resolution Accuracy
50V / 300V 0.01V / 0.1V 1% + 2dgts
Input impedance: 1 MΩ
3-2 Alternating Voltage (True RMS)
Range Resolution Accuracy(40~1KHz)
50V / 300V 0.01V / 0.1V
1.2% ± 5dgts
Input impedance: 1 MΩ
3-3 Direct Current
Range Resolution Accuracy
5000mA 1mA
50A 0.01A
1.8% + 5dgts
100A 0.01A 5% + 5 dgts