4.- INSTALLATION
4.1.- THE CB60 BATTERY CHARGER
The battery charger, CB60, has two charging compartments that can simultaneously
charge battery types BT06, BT12, BT06K and BT08K.
Connect the charger to a 120 VAC
outlet using the power source and
cable supplied. When installing the
battery charger, bear in mind that the
batteries must be charged at
temperatures over 41ºF and that the
power supply must be left on all night.
Also remember that the charger must
not be left in direct sunlight, as the
batteries will not become fully
charged at temperatures exceeding
113ºF.
Place the batteries in the charger.
The LED’s should light up, indicating that recharging is in process. Complete
recharging takes 12 hours, but the batteries may remain in the charger for an unlimited
period of time. If the LED is not illuminated, the battery has an improper connection.
Remove the battery from the charger and reseat until the LED is on.
The capacity of the batteries decreases with use. Their life span is estimated to be 500
recharging cycles, but this depends largely on the conditions of use, for which the
following is recommended:
Do not recharge the battery until it is completely flat. The transmitter indicates
when the battery is nearing a charge cycle.
Always charge the batteries at temperatures between 41ºF and 95ºF.
Charge the batteries at least once every six months regardless of use.
Avoid short-circuits between the battery contacts. Do not carry charged batteries
in toolboxes or next to other metal objects (keys, coins, etc.).
Always keep contacts clean.
Never leave batteries in direct sunlight.
Only use CERVIS manufactured batteries.
When the batteries are exhausted they can be disposed of or recycled
safely according to local standards.