30 DF II Series User Manual
SELECTING LOW RANGE LIMIT
Use the F1 key to select the LOW RANGE Pass-
Fail Limit. The cursor arrow will automatically be
placed next to the LOW Limit tag indicating that
you are about to adjust this value.
Use the UP and DOWN navigation key to increase
or decrease the Low Pass-Fail Limit VALUE. The
value will increase by the increment selected upon
each key press.
Select enter when you have adjusted both the High
and Low Pass-Fail Limits to your desired setpoints.
SELECT NOMINAL VALUE METHOD
Use the navigation pod UP and DOWN key to posi-
tion the cursor arrow next to the NOMINAL VALUE
method. Select ENTER.
TheNominalValuemethodusesaspeciedset-
point value and bandwidth to calculate the High and
Low setpoints. The DF gauge will automatically
calculate and display the High and Low Limits. You
only need to specify the Nominal Value and the %
Bandwidth.