EasyManua.ls Logo

Chicco Easy Touch - Troubleshooting

Chicco Easy Touch
96 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
21
Troubleshooting:
Error Message Problem Solution
Device stabilization in process. Wait until head stops ashing.
Battery is low and no more me-
asurements are possible.
Replace the battery.
Measurement before device
stabilization.
Wait until head stops ashing.
The device showing a rapid am-
bient temperature change.
Allow the thermometer to rest in a
room for at least 30 minutes at room
temperature:
10°C and 40°C(50°F~104°F).
The ambient temperature is not
within the range between 10°C
and 40°C (50°F~104°F).
Allow the thermometer to rest in a
room at least 30 minutes at room tem-
perature: 10°C and 40°C(50°F~
104°F).
Error 5~9, the system is not
functioning properly.
Unload the battery, wait for 1 minute
and repower it. If the message reappe-
ars, contact the retailer for service.
Temperature taken is higher
than 42.2°C (108.0°F) for fore-
head mode or 80°C (176°F) for
scan mode.
1) For forehead mode, please stay in
a stable environment for 5mins
and avoid the exercise and bath for
30min, then make sure the probe lens
is clean before take a new temperatu-
re measurement.
2) For scan mode, please select target
within specifications. If a mal-fun-
ction still exists, please contact the
nearest retailer.
Temperature taken is lower
than 34°C (93.2°F) for forehe-
ad mode or –22°C (-7.6°F) for
scan mode.
Device can not be powered on
to the ready stage.
Change with a new battery.