E19
■4-wire type resistance thermometer
No.
Measurement range Scale range
(℃)
Scale range
(K)
Measured accuracy
41
JPt100Ω1
-200.0 to 649.0 73.0 to 922.0
±
0.1%
FS ±
1
digit
Applicable only when measurement range is
“-100 to 100ºC,
-100 to 100
℃
:
-100 to 100
℃
:
±0.15%FS ±1digit
42
JPt100Ω2
-200.0 to 400.0 73.0 to 673.0
44
JPt100Ω4
-200.0 to 200.0 73.0 to 473.0
45
JPt100Ω5
-100.0 to 100.0 173.0 to 373.0
46
QPt100Ω1
-200.0 to 649.0 73.0 to 922.0
47
QPt100Ω2
-200.0 to 400.0 73.0 to 673.0
49
QPt100Ω4
-200.0 to 200.0 73.0 to 473.0
50
QPt100Ω5
-100.0 to 100.0 173.0 to 373.0
51 Pt50Ω
-200.0 to 649.0 73.0 to 922.0
52 Pt-Co
4.0 to 374.0 K 4.0 to 374.0 ±
0.15%
FS ±
1
digit
4 to 20K
:
±0.5%FS ±1digit
20 to 50K
:
±0.3%FS ±1digit
53
Pt100Ω1
-200.0 to 850.0 73.0 to1123.0
±
0.1%
FS ±
1
digit
Applicable only when measurement range is
“-100 to 100ºC,
-100 to 100
℃
:
-100 to 100
℃
:
±0.15%FS ±1digit
54
Pt100Ω2
-200.0 to 400.0 73.0 to 673.0
56
Pt100Ω4
-200.0 to 200.0 73.0 to 473.0
57
Pt100Ω5
-100.0 to 100.0 173.0 to 373.0
Input specifications(Excluding KP3000)
Thermocouple B, R, S, K, E, J, T, C(WRe5-WRe26),
W-WRe26,NiMo-Ni, CR-AuFe, N, PR5-20,
PtRh40-PtRh20, PlatinelⅡ, U, L
DC voltage ±10mV, ±20mV, ±50mV, ±100mV,
±5V, ±10V
DC current 0 to 20mA
Resistance thermometer Pt100, JPt100, Old Pt100,
Pt50, Pt-Co (4-wires)
Measurement range: Thermocouple 28 types, DC voltage 6 types,
direct current 1 type,
resistance thermometer 14 types
Temperature unit: °C, K
Accuracy rating: ±0.1%± 1 digit of measurement range
For details see, 'Detailed accuracy rating
specifications'.
Reference junction compensation accuracy : ±0.5 °C
For details see, 'Reference junction
compensation accuracy'.
Sampling rate: Approximately 0.1 seconds
Resolution: Approximately 1/30000
Burnout: Restricted to thermocouple, DC voltage (less than ±50mV),
resistance thermometer (3-wire), high limit burnout is a
standard provision. During burnout, output value of output
1 and output 2 can be set optionally, and high limit alarm is
ON (during high limit burnout) however DC voltage
(±100mV or more), DC current, resistance temperature (4
wire system) are not provided.
Input impedance: Thermocouple 1MΩ or more
DC voltage 1MΩ or more
Direct current approximately 250Ω
Allowable signal source resistance:
Thermocouple Less than 100Ω
DC voltage (mV) Less than 100Ω
DC voltage (V) Less than 300Ω
Allowable wire resistance: Resistance thermometer Less than 5Ω
(Should be common for all wires)
Measurement current of resistance thermometer:
Approximately 1mA
Maximum allowable input: Thermocouple Less than ±20V
DC voltage Less than ±20V
DC voltage Less than ±30mA,
Less than ±7.5V
Resistance thermometer
Less than 500Ω,
Less than ±5V
Maximum common mode voltage: Less than 30VAC
Common mode rejection ratio: 130dB or more (50/60Hz)
Normal mode rejection ration: 50dB or more (50/60Hz)
Output signal specifications (KP3000 only)
Output signal: Analog output 4 to 20mA, 0 to 1V, 0 to 10V
Digital output RS422A, RS485
Accuracy rating: ±0.1%FS
Output update cycle: Analog output
Approximately
0.1 seconds
Digital output
Approximately
1 second
Resolution: Approximately 1/30000
Output impedance: Voltage output Approximately 10Ω
Load resistance: Current output 400Ω or less
Voltage output 50kΩ or more
Display specifications
Upper display: LED
Lower display: LCD (with back light) 108x24 dot
Control specifications(Excluding KP3000)
Control cycle: Approximately 0.1 seconds
Output format: ON-OFF pulse type, ON-OFF servo type,
Current output type, SSR drive pulse type,
voltage output type
ON-OFF pulse type: Output signal ON-OFF pulse conductive signal
Contact capacity
Resistance load
100 to 240VAC and Less than 5A
30VDC and Less than 5A
Inductive load
100 to 240VAC and Less than 2.5A
30VDC and Less than 2.5A
Smallest load
5VDC and 10mA or more
Contact protection
Build in a small type of CR device
ON-OFF servo type: Output signal ON-OFF servo inductive signal
Contact capacity of standard load
Resistance load
100 to 240VAC and Less than 5A
30VDC and Less than 5A
Inductive load
100 to 240VAC and Less than 2.5A
30VDC and Less than 2.5A
Smallest load
5VDC and 10mA or more
Contact capacity of light load
Resistance load
100 to 240VAC and Less than 20mA
30VDC and Less than 20mA
Inductive load
100 to 240VAC and Less than 20mA
30VDC and Less than 20mA
Smallest load
5VDC and 1mA or more
Contact protection
Built in small type of CR device
Specifications