EasyManua.ls Logo

Chroma 66202 - Page 63

Default Icon
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using Remote Control
Response Parameters: <CRD>, HIGH∣LOW∣AUTO
Header on: :CURRENT:SHUNT<space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[ SOURce : ] CURRent : RANGe 66201 | 66202
Description: This command sets the current range of measure.
Setting Syntax:
CURRent:RANGe<space><CRD><PMT>
Setting Parameters: <CRD>,
66201: AUTO
∣A2∣A04∣A01∣A001
66202: High shunt: AUTO
∣A20∣A8∣A2H∣A02
Low shunt: AUTO
∣A2L∣A04∣A01∣A001
Auto shunt: AUTO
∣A20∣A8∣A2H∣A02∣A2L∣A04∣A01∣A001
Query Syntax: CURRent:RANGe?<PMT>
Response Parameters: <CRD>,
66201: AUTO
∣A2∣A04∣A01∣A001
66202: AUTO
∣A20∣A8∣A2H∣A02∣A2L∣A04∣A01∣A001
Header on: :CURRENT:RANGE<space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[ SOURce : ] CURRent : INRush : TRIGger : LEVel 66202
Description: This command sets the level of trigger of inrush current in Ampere.
Setting Syntax:
CURRent:INRush:TRIGger:LEVel<space><NR1><PMT>
Setting Parameters: <NR1>, 1 ~ 10
Query Syntax: CURRent:INRush:TRIGger:LEVel?<PMT>
Response Parameters: <NR1>, 1 ~ 10
Header on: :CURRENT:INRUSH:TRIGGER:LEVEL<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
[ SOURce : ] CURRent : INRush : TRIGger : TIME 66202
Description: This command sets the time of measure of inrush current in Millisecond.
Setting Syntax:
CURRent:INRush:TRIGger:TIME<space><NR1><PMT>
Setting Parameters: <NR1>, 1 ~ 9999
Query Syntax: CURRent:INRush:TRIGger:TIME?<PMT>
Response Parameters: <NR1>, 1 ~ 9999
Header on: :CURRENT:INRUSH:TRIGGER:TIME<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
[ SOURce : ] CURRent : INRush : TRIGger : DELay 66202
Description: This command sets the delay of measure of inrush current in Millisecond.
Setting Syntax:
CURRent:INRush:TRIGger:DELay<space><NR1><PMT>
Setting Parameters: <NR1>, 0 ~ 9999
Query Syntax: CURRent:INRush:TRIGger:DELay?<PMT>
5-21