Digital Power Meter 66200 Series User’s Manual
Response Parameters: <NR1>, 0 ~ 9999
Header on: :CURRENT:INRUSH:TRIGGER:DELAY<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
[ SOURce : ] VOLTage : RANGe 66201 | 66202
Description: This command sets the voltage range of measure.
Setting Syntax:
VOLTage:RANGe<space><CRD><PMT>
Setting Parameters: <CRD>, AUTO∣V500∣V300∣V150
Query Syntax: VOLTage:RANGe?<PMT>
Response Parameters: <CRD>, AUTO∣V500∣V300∣V150
Header on: :VOLTAGE:RANGE <space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[ SOURce : ] MEASure : MODE 66201 | 66202
Description: This command sets the mode of measure.
Setting Syntax: MEASure:MODE<space><CRD><PMT>
Setting Parameters: <CRD>, WINDOW∣AVERAGE
Query Syntax: MEASure:MODE?<PMT>
Response Parameters: <CRD>, WINDOW∣AVERAGE
Header on: :MEASURE:MODE <space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[ SOURce : ] MEASure : WINDow 66201 | 66202
Description: This command sets the time of measure over which the window calculation is to
be performed.
Setting Syntax: MEASure:WINDow<space><NR2><PMT>
Setting Parameters: <NR2>, 0.5 ~ 10.0 ,resolution 0.5
Query Syntax: MEASure:WINDow?<PMT>
Response Parameters: <NR2>, 0.5 ~ 10.0
Header on: :MEASURE:WINDOW <space><NR2><RMT>
Header off: <NR2><RMT>
Example: none
[ SOURce : ] MEASure : AVERage 66201 | 66202
Description: This command sets the number of measurements over which the average
calculation is to be performed.
Setting Syntax: MEASure:AVERage<space><NR1><PMT>
Setting Parameters: <NR1>, 1 ~ 512
Query Syntax: MEASure:AVERage?<PMT>
Response Parameters: <NR1>, 1 ~ 512
Header on: :MEASURE:AVERAGE <space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
5-22