Using Remote Control
[ SOURce : ] POWer : INTegrate 66201 | 66202
Description: This command sets the time of integration in Second.
Setting Syntax:
POWer:INTegrate<space><NR1><PMT>
Setting Parameters: <NR1>, 0 ~ 9999
Query Syntax: POWer:INTegrate?<PMT>
Response Parameters: <NR1>, 0 ~ 9999
Header on: :POWER:INTEGRATE<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
[ SOURce : ] FILTer 66201 | 66202
Description: This command is used to switch the low pass filter.
Setting Syntax:
FILTer<space><CRD><PMT>
Setting Parameters: <CRD>, ON∣OFF
Query Syntax: FILTer?<PMT>
Response Parameters: <CRD>, ON∣OFF
Header on: :FILTER<space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[ SOURce : ] ENERgy : MODE 66202
Description: This command sets the mode of energy measure.
Setting Syntax:
ENERgy:MODE<space><CRD><PMT>
Setting Parameters: <CRD>, JOULE∣WHR
Query Syntax: ENERgy:MODE?<PMT>
Response Parameters: <CRD>, JOULE∣WHR
Header on: :ENERGY:MODE<space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[ SOURce : ] ENERgy : TIME 66202
Description: This command sets the time of energy measure in Second.
Setting Syntax:
ENERgy:TIME<space><NR1><PMT>
Setting Parameters: <NR1>, 0 ~ 9999
Query Syntax: ENERgy:TIME?<PMT>
Response Parameters: <NR1>, 0 ~ 9999
Header on: :ENERGY:TIME<space><NR1><RMT>
Header off: <NR1><RMT>
[ SOURce : ] THD : MODE 66202
Description: This command sets the mode of THD measure.
Setting Syntax:
THD:MODE<space><CRD><PMT>
Setting Parameters: <CRD>, FULL∣ORDER
Query Syntax: THD:MODE?<PMT>
Response Parameters: <CRD>, FULL∣ORDER
Example: none
5-23