Using Remote Control
5-25
FETCh:POWer:REACtive?<space><NR1><PMT>,
MEASure:POWer:REACtive?<PMT>,
MEASure:POWer:REACtive?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:POWER:REACTIVE<space><NR2><RMT>
:FETCH:POWER:REACTIVE<space><NR2>,<NR2>,<NR2>,<NR2
><RMT>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:POWer:DC? {<NR1>}
MEASure[:SCALar]:POWer:DC? {<NR1>}
Type: Channel-Specific.
Description: These queries return the average power. The return could be -1, -2,
-3 and <NR2>. -1: The first time integrated calculation is not
complete yet.
-2: RCE represents “range change error” when integration process
is executing.
-3: Invalid data when OVR、OCR、OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:POWer:DC?<PMT>,
FETCh:POWer:DC?<space><NR1><PMT>,
MEASure:POWer:DC?<PMT>,
MEASure:POWer:DC?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:POWER:DC<space><NR2><RMT>
:FETCH:POWER:DC<space><NR2>,<NR2>,<NR2>,<NR2><RMT>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:POWer:ENERgy? {<NR1>}
MEASure[:SCALar]:POWer:ENERgy? {<NR1>}
Type: Channel-Specific.
Description: These queries return the energy in Joule when the ENER:MODE is
JOULE or return the energy in watt hour when the ENER:MODE is
WHR. The return could be -3 or <NR2>. -3: Invalid data when
OVR、OCR、OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:POWer:ENERgy?<PMT>,
FETCh:POWer:ENERgy?<space><NR1><PMT>,
MEASure:POWer:ENERgy?<PMT>,
MEASure:POWer:ENERgy?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:POWER:ENERGY<space><NR2><RMT>