Using Remote Control
5-29
COMMUNICATE Sub-system
[COMMunicate:]ADDRess:GPIB
Type: System.
Description: This command sets the GPIB address.
Setting syntax: ADDRess:GPIB<space><NR1><PMT>
Setting parameters: <NR1>, 1 ~ 30
Query syntax: ADDRess:GPIB?<PMT>
Return parameters: <NR1>, 1 ~ 30
Header on: :ADDRESS:GPIB<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
CONFIGURE Sub-system
[CONFigure:]VOLTage:RANGe
Type: Channel-Specific or All channels.
Description: This command sets the voltage range of measure.
Setting syntax: VOLTage:RANGe<space><CRD><PMT>
VOLTage:RANGe<space><CRD>,<CRD>,<CRD>,<CRD><PMT>
Setting parameters: <CRD>, AUTO | V600 | V300 | V150 | V60 | V30 | V15
‘/’:denote no action
Query syntax: VOLTage:RANGe?<PMT>
Return parameters: <CRD>, V600 | V300 | V150 | V60 | V30 | V15
Header on: :VOLTAGE:RANGE<space><CRD>,<CRD>,<CRD>,<CRD><RMT>
Header off: <CRD>,<CRD>,<CRD>,<CRD><RMT>
Example: none
[CONFigure:]CURRent:RANGe
Type: Channel-Specific or All channels.
Description: This command sets the current range of measure.
Setting syntax: CURRent:RANGe<space><CRD><PMT>
CURRent:RANGe<space><CRD>,<CRD>,<CRD>,<CRD><PMT>
Setting parameters: <CRD>, ‘/’:denote no action
External shunt off:
Shunt AUTO: AUTO | A20 | A5 | A2 | A05 | A02 | A005 | A002 | A0005
Shunt HIGH: AUTO | A20 | A5 | A2 | A05
Shunt LOW: AUTO | A02 | A005 | A002 | A0005
External shunt on: AUTO | E01 | E005 | E0025 | E001
Query syntax: CURRent:RANGe?<PMT>
Return parameters: <CRD>,
External shunt off: A20 | A5 | A2 | A05 | A02 | A005 | A002 | A0005
External shunt on: E01 | E005 | E0025 | E001
Header on: :CURRENT:RANGE<space><CRD>,<CRD>,<CRD>
,<CRD><RMT>
Header off: <CRD>,<CRD>,<CRD>,<CRD><RMT>
Example: none