Chroma 66204 - COMMUNICATE Sub-system; CONFIGURE Sub-system; [CONFigure:]VOLTage:RANGe

Default Icon
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using Remote Control
5-29
COMMUNICATE Sub-system
[COMMunicate:]ADDRess:GPIB
Type: System.
Description: This command sets the GPIB address.
Setting syntax: ADDRess:GPIB<space><NR1><PMT>
Setting parameters: <NR1>, 1 ~ 30
Query syntax: ADDRess:GPIB?<PMT>
Return parameters: <NR1>, 1 ~ 30
Header on: :ADDRESS:GPIB<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
CONFIGURE Sub-system
[CONFigure:]VOLTage:RANGe
Type: Channel-Specific or All channels.
Description: This command sets the voltage range of measure.
Setting syntax: VOLTage:RANGe<space><CRD><PMT>
VOLTage:RANGe<space><CRD>,<CRD>,<CRD>,<CRD><PMT>
Setting parameters: <CRD>, AUTO | V600 | V300 | V150 | V60 | V30 | V15
‘/’:denote no action
Query syntax: VOLTage:RANGe?<PMT>
Return parameters: <CRD>, V600 | V300 | V150 | V60 | V30 | V15
Header on: :VOLTAGE:RANGE<space><CRD>,<CRD>,<CRD>,<CRD><RMT>
Header off: <CRD>,<CRD>,<CRD>,<CRD><RMT>
Example: none
[CONFigure:]CURRent:RANGe
Type: Channel-Specific or All channels.
Description: This command sets the current range of measure.
Setting syntax: CURRent:RANGe<space><CRD><PMT>
CURRent:RANGe<space><CRD>,<CRD>,<CRD>,<CRD><PMT>
Setting parameters: <CRD>, ‘/’:denote no action
External shunt off:
Shunt AUTO: AUTO | A20 | A5 | A2 | A05 | A02 | A005 | A002 | A0005
Shunt HIGH: AUTO | A20 | A5 | A2 | A05
Shunt LOW: AUTO | A02 | A005 | A002 | A0005
External shunt on: AUTO | E01 | E005 | E0025 | E001
Query syntax: CURRent:RANGe?<PMT>
Return parameters: <CRD>,
External shunt off: A20 | A5 | A2 | A05 | A02 | A005 | A002 | A0005
External shunt on: E01 | E005 | E0025 | E001
Header on: :CURRENT:RANGE<space><CRD>,<CRD>,<CRD>
,<CRD><RMT>
Header off: <CRD>,<CRD>,<CRD>,<CRD><RMT>
Example: none