Digital Power Meter 66203/66204 User’s Manual
5-14
corresponding bit in the Questionable Event register.
Bit Configuration of Questionable Status Register
Bit position
15 14~6 5 4 3 2 1 0
Condition
FAN --- Energy
RCE
Inrush
RCE
Integrate
RCE
OCP OCR OVR
OVR : Over voltage range.
OCR : Over current range.
OCP : Over current protection.
Integrate RCE : Range change error when integrate mode running.
Inrush RCE : Range change error when inrush mode running.
Energy RCE : Range change error when energy mode running.
FAN : Fan failure.
Setting syntax: STATus:QUEStionable:NTRansition<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 65535
Setting example: none
Query syntax: STATus:QUEStionable:NTRansition?[<space><MAX | MIN>]<PMT>
Return parameters: <NR1>, 0 ~ 65535
Header on: :STATUS:QUESTIONABLE:NTRANSITION<space><NR1><RMT>
Header off: <NR1><RMT>
Query example: none
STATus:QUEStionable:PTRansition
Description: This command makes the values of the Questionable PTR register
set or read. Register description please refer to the description of
the previous command.
Setting syntax: STATus:QUEStionable:PTRansition<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 65535
Setting example: none
Query syntax: STATus:QUEStionable:PTRansition?[<space><MAX | MIN>]<PMT>
Return parameters: <NR1>, 0 ~ 65535
Header on: :STATUS:QUESTIONABLE:PTRANSITION<space><NR1><RMT>
Header off: <NR1><RMT>
Query example: none
STATus:PRESet
Description: This command sets the Enable, PTR, and NTR register of the status
groups to their power-on value.
Setting syntax: STATus:PRESet<PMT>
Setting parameters: none
Setting example: none
Query syntax: none
Return Parameters: none
Query example: none
STATus:CHANnel:CONDition?
Type: Channel-Specific.
Description: Returns the real time channel status.
Query syntax: STATus:CHANnel:CONDition?<PMT>