Using Remote Control
5-15
Return parameters: <NR1>
Bit Configuration of Channel Status Register
Bit position
15~6 5 4 3 2 1 0
Condition
-
Energy
RCE
Inrush
RCE
Integrate
RCE
OCP OCR OVR
Bit weight
- 32 16 8 4 2 1
Query example: STAT:CHAN:COND? Return the status of the power meter.
Return example: 2048
STATus:CHANnel:EVENt?
Type: Channel-Specific.
Description: Record all channel events that have occurred since last time the
register was read, and reset the Channel Event register.
Query syntax: STATus:CHANnel:EVENt?<PMT>
Return parameters: <NR1>
Query example: STAT:CHAN:EVEN? Read and reset the Channel Event
register.
Return example: 24
STATus:CHANnel:ENABle
Type: Channel-Specific.
Description: Mask to select which bit in the Event register is allowed to be
summed into the corresponding channel bit for the Channel
Summary Event register.
Setting syntax: STATus:CHANnel:ENABle<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 65535, Unit = None
Setting example: STAT:CHAN:ENABl 24
Query syntax: STATus:CHANnel:ENABle?[<space><MAX | MIN>]<PMT>
Return parameters: <NR1>
Query example: STAT:CHAN:ENAB? Return the contents of the Status
Channel Enable register.
Return example: 24
STATus:CHANnel:PTRansition
Type: Channel-Specific.
Description: Programmable filters that determine 0-to-1 transition in the
Condition register will set the corresponding bit of the Event register.
Setting syntax: STATus:CHANnel:PTRansition<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 65535, Unit = None
Setting example: STAT:CHAN:PTR 4 Set over current bit 2 from 0-to-1.
Query syntax: STATus:CHANnel:PTRansition?[<space><MAX | MIN>]<PMT>
Return parameters: <NR1>
Query example: STAT:CHAN:PTR?
Return example: 4