Using Remote Control
5-19
:FETCH:VOLTAGE:RMS<space><NR2>,<NR2>,<NR2>,<NR2><R
MT>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:VOLTage:PEAK+? {<NR1>}
MEASure[:SCALar]:VOLTage:PEAK+? {<NR1>}
Type: Channel-Specific.
Description: These queries return the plus value of peak voltage. The return
could be -3 or <NR2>. -3: Invalid data when OVR occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:VOLTage:PEAK+?<PMT>,
FETCh:VOLTage:PEAK+?<space><NR1><PMT>,
MEASure:VOLTage:PEAK+?<PMT>,
MEASure:VOLTage:PEAK+?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:VOLTAGE:PEAK+<space><NR2><RMT>
:FETCH:VOLTAGE:PEAK+<space><NR2>,<NR2>,<NR2>,<NR2><
RMT>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:VOLTage:PEAK-? {<NR1>}
MEASure[:SCALar]:VOLTage:PEAK-? {<NR1>}
Type: Channel-Specific.
Description: These queries return the minus value of peak voltage. The return
could be -3 or <NR2>. -3: Invalid data when OVR occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:VOLTage:PEAK-?<PMT>,
FETCh:VOLTage:PEAK-?<space><NR1><PMT>,
MEASure:VOLTage:PEAK-?<PMT>,
MEASure:VOLTage:PEAK-?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:VOLTAGE:PEAK-<space><NR2><RMT>
:FETCH:VOLTAGE:PEAK-<space><NR2>,<NR2>,<NR2>,<NR2><
RMT>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:VOLTage:DC? {<NR1>}
MEASure[:SCALar]:VOLTage:DC? {<NR1>}
Type: Channel-Specific.
Description: These queries return the DC voltage. The return could be -3 or
<NR2>.

Table of Contents