Digital Power Meter 66203/66204 User’s Manual
5-22
<NR2>.
-3: Invalid data when OCR、OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:CURRent:DC?<PMT>,
FETCh:CURRent:DC?<space><NR1><PMT>,
MEASure:CURRent:DC?<PMT>,
MEASure:CURRent:DC?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:CURRENT:DC<space><NR2><RMT>
:FETCH:CURRENT:DC<space><NR2>,<NR2>,<NR2>,<NR2><RM
T>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:CURRent:INRush? {<NR1>}
MEASure[:SCALar]:CURRent:INRush? {<NR1>}
Type: Channel-Specific.
Description: These queries return the inrush current. The return could be -3 or
<NR2>.
-3: Invalid data when OCR、OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:CURRent:INRush?<PMT>,
FETCh:CURRent:INRush?<space><NR1><PMT>,
MEASure:CURRent:INRush?<PMT>,
MEASure:CURRent:INRush?<space><NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>
Header on: :FETCH:CURRENT:INRUSH<space><NR2><RMT>
:FETCH:CURRENT:INRUSH<space><NR2>,<NR2>,<NR2>,<NR2>
<RMT>
Header off: <NR2><RMT>
<NR2>,<NR2>,<NR2>,<NR2><RMT>
Example: none
FETCh[:SCALar]:CURRent:CREStfactor? {<NR1>}
MEASure[:SCALar]:CURRent:CREStfactor? {<NR1>}
Type: Channel-Specific.
Description: These queries return the crest factor of current. The return could be
-3 or <NR2>. -3: Invalid data when OCR、OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax: FETCh:CURRent:CREStfactor?<PMT>,
FETCh:CURRent:CREStfactor?<space><NR1><PMT>,
MEASure:CURRent:CREStfactor?<PMT>,
MEASure:CURRent:CREStfactor?<space>NR1><PMT>
Query parameters: <NR1>, 0 ~ 4, 0:All channel, 1~4:Channel 1 ~ Channel 4
Return parameters: <NR2>