- Eng 23 -
English
KEY TO DISPLAY ICONS
DISPLAY MARK CONDITION/CAUSE
CORRECTIVE ACTION
Appears when the cuff is
attached too loose.
Attach the cuff properly and
measure again.
(
See page 15 “ATTACHING
THE CUFF”)
Appears when the cuff is not
attached correctly and the pulse
cannot be detected.
Appears when the pressure is
300mmHg or higher.
or
Appears when the batteries have
run out.
Replace the batteries. (
See page 11 “REMOVING THE
BATTERIES”.)
Appears when the pulse rate
falls outside the measuring range
(below 39 beats/min. or above
181 beats/min.)
Attach the cuff properly again.
Take deep breaths to relax, and
measure again. (
See page
20 “TIPS FOR OBTAINING
ACCURATE MEASUREMENT”.)
The unit does not function
properly.
Contact the store where it
was purchased or your local
service center. ( See page
26 “BEFORE REQUESTING
REPAIRS OR TESTING”.)